Bringing together ultra fast scanning technologies with high sensitivity pushes the limits of what we can see. The LCMS-8060 is built for high data quality and drives confidence and consistency 24/7.
High-Sensitivity Quantitation of Verapamil in Plasma
The triple quadrupole LCMS-8060 has an advanced ion guide design to increase ion production and detection and makes a meaningful impact on high sensitivity detection.
Actual Conc.
(ng/mL) |
Calculated Conc.
(ng/mL) |
Area RSD
(%, n=3) |
Accuracy
(%) |
0.0001 |
0.000100 |
2.97 |
100.3 |
0.0005 |
0.000508 |
6.15 |
101.7 |
0.001 |
0.000942 |
5.32 |
94.3 |
0.005 |
0.00491 |
3.54 |
98.9 |
0.01 |
0.00949 |
2.94 |
95.0 |
0.05 |
0.0511 |
2.14 |
102.4 |
0.1 |
0.0996 |
1.18 |
99.8 |
0.5 |
0.522 |
0.63 |
104.5 |
1 |
1.01 |
0.26 |
100.8 |
5 |
5.28 |
0.43 |
105.6 |
10 |
10.0 |
0.60 |
100.0 |
50 |
48.8 |
0.33 |
97.8 |