 Model   AGS-100kNX  AGS-300kNX
Capacity  100kN  300kN
Loading Method Direct, high-precision, constant-rate strain control using non-backlash precision ball-screw drive
Force
Measurement
Accuracy
 
High-Precision Type
(1/500, ±0.5%)
Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating) 
Conforms to EN 10002-2 Grade 0.5, ISO 7500-1 Class 0.5, BS 1610 Class 0.5, and ASTM E4, JIS B7721 Class 0.5.*1
Within ±0.5% indicated test force (at 1/250 to 1/1 load cell rating) 
Conforms to EN 10002-2 Grade 0.5, ISO 7500-1 Class 0.5, BS 1610 Class 0.5, and ASTM E4, JIS B7721 Class 0.5.*1
 
Standard-Precision Type
(1/500, ±1%)
 Within ±1% indicated test force (at 1/500 to 1/1 load cell rating)  

Conforms to EN 10002-2 Grade 1, ISO 7500-1 Class 1, BS 1610 Class 1, and ASTM E4, JIS B7721 Class 1. *1
 Calibration  Automatic test force calibration: select tensile, compression, or tensile and compression  
 Crosshead 
  
Speed Range  0.001 to 800 mm/min (stepless)  0.001 to 500 mm/min (stepless)
Max. Return Speed  1100mm/min  550mm/min
 Crosshead Speed Accuracy *2   ±0.1%
 Crosshead Speed and Permitted Test Force   To load cell capacity across entire speed range 0.001 to 200kN:0.001 to 500mm/min(entire speed range) 200kN to 300kN:0.001 to 400mm/min/td>
 Crosshead - Table Distance
 
(Tensile stroke) *3
1255mm (745mm, MWG)  1475mm (635mm, MWG)
 Effective Test Width   600mm  
Crosshead
Position
Detection
 Measurement Method   Optical encoder  
 Display Method   Digital display (display resolution: 0.001 mm)  
Positional Accuracy   ±0.1% indicated value or ±0.01 mm, whichever is larger 
 Sampling Speed    1000Hz max. *4  
Test Method Files 40 files (PC link: 20 files, standalone controller: 20 files)
Standard Functions
• Automatic reading of load cell characteristic values
• Test force display, stress display, stroke display,position display
• External analog output (2 channels)
• External analog input (2 channels) *4
• External digital input (2 channels) *4
• Analog recorder (option) output
• Dataletty (option) output *5
• Automatic test force / stress control (Autotuning)
• Automatic strain control (Autotuning) *4
• Test force auto-zeroing
• Test force auto-calibration
• Break detection, auto-return
• Load cell overload detection
• Touch-load detection function
 Accessories 
Load cell (with CAL cable), Power cable (5 m), turning rod, cable clamps, instruction manual 
 Dimensions
W945×D725×H2164mm
W945×D725×H2414mm
 Weight    525kg    675kg
 
Model AGS-100kNX AGS-300kNX
Power Requirements 400V model 200V model 400V model 200V model
Three phases AC380-440V
50/60Hz 4.5kVA
A three phase AC200-230V
50/60Hz 6.5kVA
Three phases 
AC380-440V
50/60Hz 5.5kVA
A three phase 
AC200-230V
50/60Hz 7.5kVA
Supply voltage fluctuations within ±10% of the set value.
C-class (10 Ω max.)
grounding resistance.
D-class (100 Ω max.)
grounding resistance.
C-class (10 Ω max.)
grounding resistance.
D-class (100 Ω max.)
grounding resistance.
Operating Environment Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation)
Floor vibrations: frequency 10 Hz max., amplitude 5 μm max.
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation.
 

This page may contain references to products that are not available in your country.
Please contact us to check the availability of these products in your country.

For Research Use Only. Not for use in diagnostic procedures.

This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.