TOC is used in a variety of applications from management of wastewater treatment plant influent and effluent, to drinking water supply management and monitoring of impurities in process and natural waters. And the range of applications in which this type of on-line technology is used is expected to continue expanding.
| Measurement type | NPOC(acidify/sparge removal of IC) and TC (standard). NPOC, TOC (TC-IC) (option). NPOC,TOC (TC-IC and POC + NPOC) (option) |
| Measurement principle | Catalytically aided combustion with NDIR detection |
| Measurement range | Variable from 0-5 ppm full scale to 0-1000ppm full scale (0-20,000 ppm is possible with on-board dilution function) |
| Reproducibility | Within +/- 2 % of full scale |
| Span stability | Within +/- 2 % of full scale/day (with temperature fluctuation within 5 °C) |
| Zero stability | Within +/- 2 % of full scale/week |
| Measurement cycle | Minimum 4 minutes (NPOC) |
| Sample dilution | Dilution with syringe, variable sample dilution factor of 2-20 |
| Auto calibration | 1-2 point auto calibration possible |
| Multi-stream option | 2 streams with optional sample switching set, Maximum 6 streams with optional flow line switcher |
| Display | LCD 40 characters x 14 lines |
| Recorder option | Recorder width is 100 mm, 6 dot recorder |
| Printer option | Thermal type, 40 columns, chart width with 110 mm. |
| Analog output | Selectable output of 0-1 VDC, 0-16 mA or 4-20 mA (insulated output) |
| Output | RS 422 |
| Alarm | Selectable among upper/lower limit, upper/upper limit and lower/lower limit and instrument trouble alarm |
| Input signal | Sample measurement start, sample measurement stop, calibration start, alarm reset |
| Event signals | Ready state, operating online, measuring, calibrating and more. |
| Carrier gas | Air or oxygen, supply pressure: 300 kPa, free of dust, oil mist, water droplets |
| Power | 100,110,115,120,220,230,240 V as ordered; 5A; 50-60 Hz |
| Construction | Indoor wall mounting (optional stand available) |