Product

Laboratory Instruments

TOC-4110

TOC Analyzers

TOC-4110

TOC is used in a variety of applications from management of wastewater treatment plant influent and effluent, to drinking water supply management and monitoring of impurities in process and natural waters. And the range of applications in which this type of on-line technology is used is expected to continue expanding.

Features

  • High-performance on-line TOC analyzer using the established 680°C catalyst-aided combustion and non-dispersive infrared detection methods.
  • Wide measurement range is variable from 0-5ppm full scale to 0-1000ppm full scale. Measurements up to 20000ppm are possible using the on-board dilution function.
  • Measurement mode selectable between TC/NPOC (standard), TOC=TC-IC (optional) or TOC=POC+NPOC.

Specifications

Measurement type NPOC(acidify/sparge removal of IC) and TC (standard). NPOC, TOC (TC-IC) (option). NPOC,TOC (TC-IC and POC + NPOC) (option)
Measurement principle Catalytically aided combustion with NDIR detection
Measurement range Variable from 0-5 ppm full scale to 0-1000ppm full scale (0-20,000 ppm is possible with on-board dilution function)
Reproducibility Within +/- 2 % of full scale
Span stability Within +/- 2 % of full scale/day (with temperature fluctuation within 5 °C)
Zero stability Within +/- 2 % of full scale/week
Measurement cycle Minimum 4 minutes (NPOC)
Sample dilution Dilution with syringe, variable sample dilution factor of 2-20
Auto calibration 1-2 point auto calibration possible
Multi-stream option 2 streams with optional sample switching set, Maximum 6 streams with optional flow line switcher
Display LCD 40 characters x 14 lines
Recorder option Recorder width is 100 mm, 6 dot recorder
Printer option Thermal type, 40 columns, chart width with 110 mm.
Analog output Selectable output of 0-1 VDC, 0-16 mA or 4-20 mA (insulated output)
Output RS 422
Alarm Selectable among upper/lower limit, upper/upper limit and lower/lower limit and instrument trouble alarm
Input signal Sample measurement start, sample measurement stop, calibration start, alarm reset
Event signals Ready state, operating online, measuring, calibrating and more.
Carrier gas Air or oxygen, supply pressure: 300 kPa, free of dust, oil mist, water droplets
Power 100,110,115,120,220,230,240 V as ordered; 5A; 50-60 Hz
Construction Indoor wall mounting (optional stand available)