Product

Laboratory Instruments

TN-4110

TOC Analyzers

TN-4110

TOC is used in a variety of applications from management of wastewater treatment plant influent and effluent, to drinking water supply management and monitoring of impurities in process and natural waters. And the range of applications in which this type of on-line technology is used is expected to continue expanding. In addition, because nitrogen is considered to be one of the factors which contributes to the over-proliferation of organisms in enclosed bodies of water, management of these public water sources and of treatment plant influent and effluent also includes demand for monitoring of nitrogen levels.

Features

  • The TN-4110 is high-performance TN analyzer that uses the combustion/chemiluminescence detection technique.
  • The TN measurement range is from 0-5ppm full scale to 0-200ppm full scale with the dilution function up to 4000ppm.
  • With the short 4-minute measurement cycle, sudden changes in TN concentration can be caught.

Specifications

Measurement type Total Nitrogen TN
Measurement principle Combustion decomposition with chemiluminescence detection
Measurement range Variable from 0-1 ppm full scale to 0-200 ppm full scale (0-4000 ppm is possible with on-board dilution function)
Reproducibility Within +/- 4 % of full scale at less than 4 ppm full scale, within +/- 2 % full scale at greater then 4 ppm full scale
Span stability Within +/- 2 % of full scale/day (with temperature fluctuation within 5 °C)
Zero stability Within +/- 2 % of full scale/week
Measurement cycle Minimum 4 minutes
Sample dilution Dilution with syringe, variable sample dilution factor of 2-20
Auto calibration 1-2 point auto calibration possible
Multi-stream option 2 streams with optional sample switching set, Maximum 6 streams with optional flow line switcher
Display LCD 40 characters x 14 lines
Recorder option Recorder width is 100 mm, 6 dot recorder
Printer option Thermal type, 40 columns, chart width with 110 mm.
Analog output Selectable output of 0-1 VDC, 0-16 mA or 4-20 mA (insulated output)
Output RS 422
Alarm Selectable among upper/lower limit, upper/upper limit and lower/lower limit and instrument trouble alarm
Input signal Sample measurement start, sample measurement stop, calibration start, alarm reset
Event signals Ready state, operating online, measuring, calibrating and more.
Carrier gas Air or oxygen, supply pressure: 300 kPa, free of dust, oil mist, water droplets
Power 100,110,115,120,220,230,240 V as ordered; 5A ;50-60 Hz
Construction Indoor wall mounting (optional stand available)