Installation

1. Installation room environmen

Temperature: 18 to 28(rate of temperature change: 2 degrees Celsius /h or lower)
Humidity: 20 to 70%
Avoid locations with high levels of vibration or dust.
 

2. Power source

Main unit: Single-phase 200/220/230/240V 10% (max 250V) 50/60Hz 30A
Autosampler ASC-6100F: Single-phase 100-240V 10% 50/60Hz 1A (Option)
Ultrasonic nebulizer UAG-1: Single-phase 100V 10% 50/60Hz 10A (Option)
Hydride generator HVG-ICP: Single-phase 100V 10% 50/60Hz 2A (Option)
* All of the above are grounded and require 3-prong sockets.
** Check separately when using the cooling water circulator.
 

3. Grounding

Resistance: independent grounding below 30 ohms
 

4. Gas installation

Type: Argon gas of 99.95% or greater purity Advjust so that argon gas supply pressure is 45010 kPa. Approximately one 7-cubic meter gas cylinder is required per 8 hours of operation. Consumption rate of argon gas used in analysis is 0.74 cubic meters/h with the standard specifications.
 

5. Cooling water

Plasma stand cooling water: Water temperature 5 to 30 flow rate 1L/min or more CCD detector cooling water: Use Shimadzu's recommended cooler Note) UAG-1 requires separate cooling water supply. Note) Cooling water circulator can also be used for plasma stand and UAG-1.
 

6. Exhaust duct

Plasma stand and radio frequency power source exhaust duct Exhaust gas from the plasma stand is primarily argon, however, be sure to install an exhaust duct, since this gas may also contain metallic vapors and solvents
 
 